An evaluation of the test-retest reliability of the functional head impulse test in healthy young adults

Küçük Resim Yok

Tarih

2023

Dergi Başlığı

Dergi ISSN

Cilt Başlığı

Yayıncı

Springer

Erişim Hakkı

info:eu-repo/semantics/closedAccess

Özet

AimTo determine the test and retest reliability of the functional head impulse test (fHIT) in healthy young adults.Materials and methodsThirty-three healthy participants (17 women, 16 men) aged 18-30 years were included in the study. Each participant underwent the fHIT twice, 1 week apart, by the same experienced clinician. Intraclass correlation coefficients (ICCs) were used to determine test-retest reliability.ResultsThere was no statistically significant difference between the results of total percentage of correct answer (CA%) of the fHIT obtained in session 1 and session 2 measurements in the lateral, anterior, and posterior semicircular canals (SCCs) (p > 0.05). ICC values for test-retest reliability were found to range from 0.619 to 0.665 for the three semicircular canals (SCCs).ConclusionThe test-retest reliability of the fHIT device was moderate. Attention, cognition, and fatigue may be the factors reducing reliability. In the diagnosis, follow-up, and rehabilitation processes of vestibular diseases in clinics, changes in the fHIT CA% can be used to assess vestibulo-ocular reflex (VOR) functionality.

Açıklama

Anahtar Kelimeler

fHIT, Vestibulo-ocular reflex, Test-retest reliability, Intraclass correlation coefficient

Kaynak

European Archives of Oto-Rhino-Laryngology

WoS Q Değeri

Q2

Scopus Q Değeri

Cilt

280

Sayı

12

Künye