Numerical Analysis of Spinal Cord Stimulation with an Eight-Shape Electrode Model
Küçük Resim Yok
Tarih
2022
Yazarlar
Dergi Başlığı
Dergi ISSN
Cilt Başlığı
Yayıncı
Ieee
Erişim Hakkı
info:eu-repo/semantics/closedAccess
Özet
Spinal Cord Stimulation is used for drug-free treatment in indications of chronic pain of the spinal cord. As a result of sending current from the electrodes placed in the epidural region of the spinal cord with laminectomy, pain information is prevented from reaching the brain and the patient only feels paresthesia. Today, different shapes of electrode designs are available. In this study, the current density distribution resulting from the application of 0.0417 mA to the anode and -0.0417 mA to the cathode from the eight-shaped electrodes were examined. The maximum current density is 36 mA/cm(2), and this value is within the safety limits that the patient can feel.
Açıklama
Medical Technologies Congress (TIPTEKNO) -- OCT 31-NOV 02, 2022 -- Antalya, TURKEY
Anahtar Kelimeler
Spinal cord stimulation, chronic pain, numerical analysis, electrode
Kaynak
2022 Medical Technologies Congress (Tiptekno'22)
WoS Q Değeri
N/A
Scopus Q Değeri
N/A