Numerical Analysis of Spinal Cord Stimulation with an Eight-Shape Electrode Model

Küçük Resim Yok

Tarih

2022

Dergi Başlığı

Dergi ISSN

Cilt Başlığı

Yayıncı

Ieee

Erişim Hakkı

info:eu-repo/semantics/closedAccess

Özet

Spinal Cord Stimulation is used for drug-free treatment in indications of chronic pain of the spinal cord. As a result of sending current from the electrodes placed in the epidural region of the spinal cord with laminectomy, pain information is prevented from reaching the brain and the patient only feels paresthesia. Today, different shapes of electrode designs are available. In this study, the current density distribution resulting from the application of 0.0417 mA to the anode and -0.0417 mA to the cathode from the eight-shaped electrodes were examined. The maximum current density is 36 mA/cm(2), and this value is within the safety limits that the patient can feel.

Açıklama

Medical Technologies Congress (TIPTEKNO) -- OCT 31-NOV 02, 2022 -- Antalya, TURKEY

Anahtar Kelimeler

Spinal cord stimulation, chronic pain, numerical analysis, electrode

Kaynak

2022 Medical Technologies Congress (Tiptekno'22)

WoS Q Değeri

N/A

Scopus Q Değeri

N/A

Cilt

Sayı

Künye