The effect of the cochlear implant electrode position on the neural response telemetry results: A prospective clinical trial at a single center
dc.contributor.author | Deniz, Huseyin | |
dc.contributor.author | Yazici, Alper | |
dc.date.accessioned | 2022-03-08T07:15:24Z | |
dc.date.available | 2022-03-08T07:15:24Z | |
dc.date.issued | 2019 | |
dc.department | İnönü Üniversitesi | en_US |
dc.description.abstract | Aim: This study aims to investigate the changes in the post-op NRT thresholds of two different implant electrodes of the same brand positioned laterally within the scala tympani, and the pre-modiolar locations, using the same round window approach.Material and Methods: After CI operation NRT measurements conducted at different times on the Nucleus® CI422 nucleus slim straight and Nucleus® CI24RE electrode types of two different electrode models.Results: The number of patients who received the Nucleus® CI24RE (ST) electrode and the Nucleus® CI422 nucleus slim straight were 26 (33.3%) and 56 (66.7%), respectively. NRT values tended to decrease from the basal part of cochlea towards an apex at any time of evaluation in both types of cochlear implants. Conclusion: There was a difference between Nucleus® CI24RE and Nucleus® CI422 nucleus slim straight groups in terms of NRT thresholds. Nucleus® CI422 nucleus slim straight electrode yielded lower NRT threshold levels. | en_US |
dc.identifier.citation | Deniz, H., & Yazici, A. (2021). The effect of the cochlear implant electrode position on the neural response telemetry results: A prospective clinical trial at a single center . Annals of Medical Research, | en_US |
dc.identifier.uri | https://hdl.handle.net/11616/54633 | |
dc.language.iso | en | en_US |
dc.relation.ispartof | Annals of Medical Research | en_US |
dc.rights | info:eu-repo/semantics/openAccess | en_US |
dc.title | The effect of the cochlear implant electrode position on the neural response telemetry results: A prospective clinical trial at a single center | en_US |
dc.type | Article | en_US |