Angulated versus parallel headless cannulated screw fixation for type 1 capitellum fracture: A finite element study

dc.contributor.authorEfendigil, Tugba
dc.contributor.authorAlp, Nazmi Bulent
dc.contributor.authorDemir, Ozgur
dc.contributor.authorBilsel, Kerem
dc.contributor.authorKalay, Onur Can
dc.contributor.authorKarpat, Fatih
dc.date.accessioned2022-03-14T08:01:10Z
dc.date.available2022-03-14T08:01:10Z
dc.date.issued2020
dc.departmentİnönü Üniversitesien_US
dc.description.abstractAim: The unique shape and vascularization make capitellar fractures challenging to treat. In such cases, screw inclination is crucial to achieve stable fixation. In the present study, we aimed to evaluate the mechanical outcome of the different fixation angles formed by headless cannulated screws for treatment of type 1 capitellum fracture and compared interfragmentary displacement (IFD) using finite element analysis (FEA). Material and Methods: In our study, three-dimensional finite element stress analysis was applied using the isotropic materials and static linear analysis. Stochastic screw inclination scenarios (Case 1, Case 2, Case 3, and Case 4) were generated, and perpendicular application of loads were simulated with magnitudes between 50 and 300 N on the elbow at angulations from 0º to 145º. The IFD in the four different screw inclination pairs were listed in the ANSYS general end processor. Relative IFD was calculated by measuring the total displacements in the X/Y/Z planes from 16 different points.Results: According to the modeling, the magnitude of IFD was significantly different in terms of the force and the case factors under four altered forces. Analysis of displacement exhibited a significant difference in the force values of ‘100N/200N/300N’ and in the screw inclinations of Case 2 only. Although the interaction effect between the Force*Case was observed to be low, the most noticeable results were obtained from the “300N*Case 2.” pair compared to the others (p0.05).Conclusion: Screws used for fracture fixation must maintain an anatomical correction, until it heals. Following a complex trauma such as a capitellar fracture, the elbow joint needs a stable fixation and early mobilization. Our FEA results suggest that the fixation via angled screws can achieve a more stable configuration than the paralleled ones.en_US
dc.identifier.citationBulent Alp, N., Efendigil, T., Demir, O., Bilsel, K., Can Kalay, O., & Karpat, F. (2021). Angulated versus parallel headless cannulated screw fixation for type 1 capitellum fracture: A finite element study . Annals of Medical Research,en_US
dc.identifier.urihttps://hdl.handle.net/11616/55223
dc.language.isoenen_US
dc.relation.ispartofAnnals of Medical Researchen_US
dc.rightsinfo:eu-repo/semantics/openAccessen_US
dc.titleAngulated versus parallel headless cannulated screw fixation for type 1 capitellum fracture: A finite element studyen_US
dc.typeArticleen_US

Dosyalar

Orijinal paket
Listeleniyor 1 - 1 / 1
Yükleniyor...
Küçük Resim
İsim:
2550-2555.pdf
Boyut:
1.27 MB
Biçim:
Adobe Portable Document Format
Açıklama:
Lisans paketi
Listeleniyor 1 - 1 / 1
Küçük Resim Yok
İsim:
license.txt
Boyut:
1.71 KB
Biçim:
Item-specific license agreed upon to submission
Açıklama: