Evaluation of Metal Artifacts in Cone Beam Computed Tomography by Metal Supported Porcelain Crowns Using Different FOV and Localizations: An In Vitro Study

dc.authorscopusid55347581500
dc.authorscopusid55800550000
dc.contributor.authorDedeoğlu N.
dc.contributor.authorYaşa Y.
dc.date.accessioned2024-08-04T20:03:39Z
dc.date.available2024-08-04T20:03:39Z
dc.date.issued2020
dc.departmentİnönü Üniversitesien_US
dc.description.abstractIntroduction: Metal-supported porcelain crowns (MSPC) and bridge restorations may be present in the mouths of patients undergoing CBCT imaging. Artifacts that are caused by these MSPCs may adversely affect image quality. The aim of this study is to determine the effect of different FOV (field of view) and localization in FOV on metal artifacts caused by MSPC. Methods: Twenty MSPCs scanned with CBCT at central and peripheral localization at 18x16 and 8x8 cm FOV. The 2.5 mm periphery area of the MSPC cross-sectional image was evaluated. The metal artefact-area within this area was measured. Then, the artifact-area to total-area ratio was converted to form of a percentage. In addition to evaluation of crown periphery area, the lengths of the metal streaks artifacts were measured. The maximum linear dimension of the metal artifact was recorded from the crown margin for each MSPC in cross-sectional image in the bucco-lingual direction. All data collected were evaluated by Kruskal-Wallis analysis and Mann Whitney U test (P<0,05). Results: No statistically significant differences were found in the artifacts-area measured (P=0.121). However, there was a statistically significant difference in linear dimension measurements of artifacts (P=0.000). In 18x16 cm FOV localization peripheral linear dimension measurements were higher than other FOV and localizations. Conclusion: Linear size artifacts of MSPC were found to be higher in peripheral positioning in wide FOV. However, according to this study, areas evaluated for metal artifacts caused by MSPCs are not affected by FOV and localization. © 2020, Mashhad University of Medical Sciences. All rights reserved.en_US
dc.identifier.doi10.22038/JDMT.2020.48334.1369
dc.identifier.endpage87en_US
dc.identifier.issn2322-4150
dc.identifier.issue2en_US
dc.identifier.scopus2-s2.0-85186431687en_US
dc.identifier.scopusqualityN/Aen_US
dc.identifier.startpage81en_US
dc.identifier.urihttps://doi.org/10.22038/JDMT.2020.48334.1369
dc.identifier.urihttps://hdl.handle.net/11616/91985
dc.identifier.volume9en_US
dc.indekslendigikaynakScopusen_US
dc.language.isoenen_US
dc.publisherMashhad University of Medical Sciencesen_US
dc.relation.ispartofJournal of Dental Materials and Techniquesen_US
dc.relation.publicationcategoryMakale - Uluslararası Hakemli Dergi - Kurum Öğretim Elemanıen_US
dc.rightsinfo:eu-repo/semantics/closedAccessen_US
dc.subjectCBCTen_US
dc.subjectDental image qualityen_US
dc.subjectFOVen_US
dc.subjectIn-vitro studyen_US
dc.subjectMetal artifacten_US
dc.subjectMetal crownen_US
dc.titleEvaluation of Metal Artifacts in Cone Beam Computed Tomography by Metal Supported Porcelain Crowns Using Different FOV and Localizations: An In Vitro Studyen_US
dc.typeArticleen_US

Dosyalar