High-sensitivity electrochemical sensor using no nanomaterials for the detection of ciprofloxacin with poly 2-(hydroxymethyl)thiophene-modified glassy carbon electrode
Küçük Resim Yok
Tarih
2024
Dergi Başlığı
Dergi ISSN
Cilt Başlığı
Yayıncı
Springer
Erişim Hakkı
info:eu-repo/semantics/closedAccess
Özet
In this study, a new sensor was developed for the detection of ciprofloxacin that is easy to prepare, selective and has excellent sensitivity through electrochemical polymerization of 2-(hydroxymethyl)thiophene on glassy carbon (GC) electrode. Electrode modifications were performed by using cyclic voltammetry. The morphology and surface properties of the modified electrode were characterized by using scanning electron microscopy. Determination of ciprofloxacin with poly 2-(hydroxymethyl)thiophene-modified GC electrode was performed by square wave voltammetry in 0.1 M citrate buffer (pH 5.00) at 1.1 V. The prepared sensor showed a wide linear range from 0.1 to 200 mu M, detection limit of 7 nM, a limit of quantitation of 22 nM, a correlation coefficient of 0.9993, excellent sensitivity of 99.6% and a relative standard deviation of 0.40%. The sensor was used to determine ciprofloxacin in human urine samples, and a recovery efficiency of 99.24-100.36% was obtained. According to the results in this study, we believe that the sensor developed can be used for the detection of ciprofloxacin in clinical laboratories.
Açıklama
Anahtar Kelimeler
Ciprofloxacin, Electrochemical sensor, Antibiotic, 2-(Hydroxymethyl)thiophene polymer, Voltammetry
Kaynak
Polymer Bulletin
WoS Q Değeri
Q2
Scopus Q Değeri
Q2
Cilt
81
Sayı
7