Determination of antimony by using tungsten trap atomic absorption spectrometry

Küçük Resim Yok

Tarih

2008

Dergi Başlığı

Dergi ISSN

Cilt Başlığı

Yayıncı

Pergamon-Elsevier Science Ltd

Erişim Hakkı

info:eu-repo/semantics/closedAccess

Özet

An electrically heated tungsten coil was used as a trap in the determination of antimony. The technique consists of three steps. Initially, SbH(3) is formed by hydride generation procedure; then the analyte species in vapor form are transported to W-coil trap heated at 370 degrees C. Following the preconcentration step, the trap is heated to 895 degrees C; analyte species are revolatilized and transported to the flame-heated quartz atom cell where atornization and the formation of signal take place. The experimental parameters were optimized both for trap and no-trap studies. The most important experimental parameters are concentrations of HCI and NaBH4 solutions. H2 and At gas flow rates. and collection and revolatilization temperatures of W-coil. Accuracy was tested using a certified reference material, waste water EU-L-1. Limit of detection for the system is 16 ng 1- 1 using a sample of 36 ml collected in 4.0 min. Enhancement factor in sensitivity was 17. (c) 2008 Elsevier B.V. All rights reserved.

Açıklama

35th Biennial Colloquium Spectroscopicum Internationale -- SEP 23-27, 2007 -- Xiamen, PEOPLES R CHINA

Anahtar Kelimeler

antimony, hydride generation, W-trap, atomic absorption spectrometry, in-situ preconcentration

Kaynak

Spectrochimica Acta Part B-Atomic Spectroscopy

WoS Q Değeri

Q1

Scopus Q Değeri

Q2

Cilt

63

Sayı

8