Evaluation of trueness and precision of 3 face-scanning devices

Küçük Resim Yok

Tarih

2025

Dergi Başlığı

Dergi ISSN

Cilt Başlığı

Yayıncı

Mosby-Elsevier

Erişim Hakkı

info:eu-repo/semantics/closedAccess

Özet

Introduction: This study evaluated the trueness and precision of three 3-dimensional facial scanning devices- MetiSmile (MS), Revopoint Miraco Plus (RMP), and Qlone Dental application (QA) using an iPhone 15 Pro. Methods: Forty-seven participants were scanned using MS, RMP, and QA under standardized conditions. Comprehensive accuracy evaluations (trueness and precision) were conducted. Deviations were assessed through superimposition, and the root mean square (RMS) was used to quantify these deviations. The deviation analysis was performed using CloudCompare software (v2.13.2, www.cloudcompare.org). Results: Trueness analysis showed that MS and RMP produced comparable results with deviations below 0.5 mm, whereas QA exhibited significantly higher deviations (P = 0.003; mean RMS = 0.938 mm for full-head scans). Precision analysis revealed MS had the lowest mean RMS deviation (0.309 mm), indicating the highest consistency, followed by RMP (0.362 mm). QA demonstrated the least precision (mean RMS = 0.657 mm). Comparative analysis indicated MS and RMP scans were similar, whereas QA showed greater deviations. Conclusions: The MS and RMP devices exhibit high accuracy, making them suitable for orthodontic diagnostics and treatment planning. Although less accurate, the QA is also applicable for these purposes. (Am J Orthod Dentofacial Orthop 2025;168:358-66)

Açıklama

Anahtar Kelimeler

Accuracy

Kaynak

American Journal of Orthodontics and Dentofacial Orthopedics

WoS Q Değeri

Q1

Scopus Q Değeri

Q1

Cilt

168

Sayı

3

Künye